LH 500 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2014-01457
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- January 17, 2014
- Report Date
- January 17, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K032000
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. THE FSE NOTICED A LEAK OF CLEANER INSIDE THE INSTRUMENT AND DISCOVERED A DEFECTIVE TUBING AT PINCH VALVE PV37. THE FSE REPLACED THE TUBING AT PV37 TO RESOLVE THE LEAK. THE FSE ALSO FOUND A SECOND ISSUE ASSOCIATED WITH A LEAK OF DILUENT INSIDE THE INSTRUMENT. THE FSE REPLACED THE TUBING FOR PINCH VALVE PV49 TO RESOLVE THE SECOND LEAK. THE FSE THEN DISCOVERED A THIRD LEAK ASSOCIATED WITH THE PROBE WIPE LEAKING ON THE FRONT OF THE ANALYZER. THE FSE FOUND THE PROBE WAS BENT THEREBY NOT ALLOWING THE PROBE WIPE TO COMPLETE THE FULL LENGTH OF TRAVEL TO RINSE THE PROBE. IN ADDITION, THERE WAS AN INCORRECTLY PLACED TUBING ON PINCH VALVE PV42 CAUSING EXCESSIVE DILUENT TO LEAK FROM THE PROBE RINSE BLOCK. THE FSE STRAIGHTENED THE BENT PROBE AND REPLACED THE INCORRECTLY PLACED TUBING AT PV42 TO RESOLVE THE LEAK FROM THE PROBE. THE FSE WAS THEN INFORMED BY THE CUSTOMER THAT THE INSTRUMENT GENERATED RED BLOOD CELL (RBC) AND MEAN CORPUSCULAR VOLUME (MCV) VOTE OUTS (NON-NUMERICAL RESULTS). THE FSE PROCEEDED TO REPLACE THE O-RINGS AND RBC APERTURE TO RESOLVE THE RBC/MCV VOTE OUTS. ERRONEOUS PATIENT RESULTS WERE NOT GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT. THERE WAS NO REPORT OF INJURY OR DIRECT EXPOSURE TO THE LEAK IN CONNECTION WITH THIS EVENT. (B)(4).
THE CUSTOMER REPORTED OBTAINING A BIASED LOW RED BLOOD CELL (RBC) PARAMETER FROM THE LH 500 HEMATOLOGY ANALYZER FOR THE COULTER 5C CELL CONTROL WHEN PERFORMING CALIBRATION WITH THE COULTER S-CAL CALIBRATOR KIT, AS COMPARED TO THE RBC ASSAY VALUE ON THE COULTER 5C CELL CONTROL. ERRONEOUS PATIENT RESULTS WERE NOT GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. CUSTOMER TECHNICAL SPECIALIST (CTS) ASSISTED THE CUSTOMER WITH TROUBLESHOOTING OVER THE TELEPHONE AND NOTED THE RBC CALIBRATION FACTOR HAS BEEN CALCULATED AND ADJUSTED TO RECOVER WITHIN THE ASSIGNED RANGE FOR THE 5C CONTROLS. THE CUSTOMER REQUESTED FOR SERVICE TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379935 | LH 500 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |