FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 3900952 · Received June 27, 2014

Report

Report Number
3004209178-2014-86317
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE PRIME TEST DUE TO A FAULTY FORCE SENSOR RESISTOR. THE DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INFUSION SET AND RESERVOIR WERE CHANGED AND DID A REWIND THREE TO FOUR TIMES WITH A COMPLETE INFUSION SET AND BATTERY CHANGE. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS LOOSE. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377698 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-715WWL

Patients

Seq Age Sex Outcome Treatment
1