FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3900930
·
Received June 27, 2014
Report
- Report Number
- 3004209178-2014-86307
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- May 10, 2014
- Report Date
- June 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS. UNABLE TO PRIME THE UNIT DURING THE PRIME TEST AS A RESULT OF A LOOSE/FLUSH DRIVE SUPPORT. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS, SCRATCHED SCREEN, AND MISSING END CAP STICKER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE READING WAS 145MG/DL. THE CALLER STATED THAT THE DEVICE WENT TO REWIND AND THEN THE ALARM WENT OFF. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 376742 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |