FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3900929 · Received June 27, 2014

Report

Report Number
3004209178-2014-86306
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
May 6, 2014
Report Date
June 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE DEVICE HAD SCRATCHED SCREEN, CRACKED DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED. NO FURTHER INFORMATION WAS PROVIDED.