FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3900924
·
Received June 27, 2014
Report
- Report Number
- 2032227-2014-03400
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- March 20, 2014
- Report Date
- June 24, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH ALARMS DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, RESERVOIR TUBE LIP AND SCRATCHED DISPLAY WINDOW.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT HE HAD HIGH BLOOD GLUCOSE OF 516 MG/DL AND THE INSULIN PUMP WAS NOT WORKING CORRECTLY. CUSTOMER STATED THAT THE INSULIN PUMP WAS ALARMING AND SQUIRTING THE INSULIN OUT DURING PRIME REWIND PROCESS. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 376928 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |