FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3900924 · Received June 27, 2014

Report

Report Number
2032227-2014-03400
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
March 20, 2014
Report Date
June 24, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH ALARMS DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, RESERVOIR TUBE LIP AND SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HE HAD HIGH BLOOD GLUCOSE OF 516 MG/DL AND THE INSULIN PUMP WAS NOT WORKING CORRECTLY. CUSTOMER STATED THAT THE INSULIN PUMP WAS ALARMING AND SQUIRTING THE INSULIN OUT DURING PRIME REWIND PROCESS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376928 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAB

Patients

Seq Age Sex Outcome Treatment
1 19 YR