FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3900894 · Received June 27, 2014

Report

Report Number
2032227-2014-03401
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
April 22, 2014
Report Date
June 6, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST WITH UNITS REMAINING ON THE STATUS SCREEN. UNIT HAD MOTOR ERROR ALARMS DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNIT HAD CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS, SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP WAS ALARMING MOTOR ERROR AND CUSTOMER WAS UNABLE TO CLEAR IT. CALLER STATED THAT THE CUSTOMER HAD AN MRI ON THE KNEE AND THE INSULIN PUMP WAS EXPOSED TO A MAGNETIC FIELD. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377067 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754LCMS

Patients

Seq Age Sex Outcome Treatment
1