FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3900894
·
Received June 27, 2014
Report
- Report Number
- 2032227-2014-03401
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- April 22, 2014
- Report Date
- June 6, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TEST WITH UNITS REMAINING ON THE STATUS SCREEN. UNIT HAD MOTOR ERROR ALARMS DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNIT HAD CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS, SCRATCHED DISPLAY WINDOW.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP WAS ALARMING MOTOR ERROR AND CUSTOMER WAS UNABLE TO CLEAR IT. CALLER STATED THAT THE CUSTOMER HAD AN MRI ON THE KNEE AND THE INSULIN PUMP WAS EXPOSED TO A MAGNETIC FIELD. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 377067 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754LCMS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |