FDA Adverse Event Malfunction Summary report: N

AUTOPULSE® RESUSCITATION SYSTEM MODEL 100

MDR report key: 3900865 · Received June 27, 2014

Report

Report Number
3003793491-2014-00295
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 4, 2014
Report Date
June 5, 2014
Manufacturer
ZOLL CIRCULATION, INC
Product Code
DRM
PMA / PMN Number
K112998
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PARAMEDIC

Narratives

Additional Manufacturer Narrative · 1

ZOLL HAS NOT YET RECEIVED THE AUTOPULSE PLATFORM IN COMPLAINT FOR INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE FILED IF AND WHEN THE PRODUCT IS RETURNED AND INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE AUTOPULSE PLATFORM IN COMPLAINT WAS RETURNED TO ZOLL ON 06/30/2014 FOR INVESTIGATION. VISUAL INSPECTION OF THE RETURNED PLATFORM WAS PERFORMED AND FOUND THE LCD BACKLIGHT TO NOT BE FUNCTIONING. INITIAL FUNCTIONAL TESTING OF THE PLATFORM WAS PERFORMED AND NO FAULTS OR ERRORS WERE EXHIBITED. A REVIEW OF THE ARCHIVE WAS UNABLE TO BE PERFORMED AS THE ARCHIVE WAS UNABLE TO BE DOWNLOADED BECAUSE THE IR AND SERIAL PORTS OF THE PROCESSOR PCA BOARD WERE DEFECTIVE. THE PROCESSOR PCA BOARD WAS REPLACED REMEDYING THE CUSTOMER'S REPORTED COMPLAINT. THE PLATFORM UNDERWENT AND MET ALL FINAL TESTING CRITERIA. THE CUSTOMER'S REPORTED COMPLAINT OF THE LCD BACKLIGHT NOT FUNCTIONING WAS CONFIRMED THROUGH VISUAL INSPECTION OF THE PLATFORM. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE PROCESSOR PCA AT THE LCD BACKLIGHT INPUT. THE PROCESSOR PCA BOARD WAS REPLACED, REMEDYING THE CUSTOMER'S REPORTED COMPLAINT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INSPECTION, IT WAS FOUND THAT THE AUTOPULSE PLATFORM'S DISPLAY WAS DAMAGED. NO INFORMATION CAN BE CLEARLY SEEN. OTHER THAN THIS, THE DEVICE WORKS FINE. NO PATIENT INVOLVEMENT WAS REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376711 AUTOPULSE® RESUSCITATION SYSTEM MODEL 100 CARDIAC CHEST COMPRESSOR DRM ZOLL CIRCULATION, INC MODEL 100

Patients

Seq Age Sex Outcome Treatment
1