FDA Adverse Event Injury Summary report: N

RSP SHOULDER

MDR report key: 3900731 · Received June 27, 2014

Report

Report Number
1644408-2014-00397
Event Type
Injury
Date Received
June 27, 2014
Date of Event
June 19, 2014
Report Date
June 19, 2014
Manufacturer
ENCORE MEDICAL, L.P.
Product Code
KWS
PMA / PMN Number
K051075
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REASON FOR THIS REVISION SURGERY WAS REPORTED AS A CHRONIC INFECTION. THE HEALTHCARE PROFESSIONAL INDICATED A SIGNIFICANT ADVERSE EVENT TO THE PATIENT. THERE WAS NO DELAY IN SURGERY AND ANOTHER SUITABLE DEVICE WAS AVAILABLE FOR USE. THE REVISION SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS DISPOSED OF AT THE HOSPITAL AND NOT MADE AVAILABLE TO DJO SURGICAL FOR EXAMINATION. A REVIEW OF THE IMPLANT DEVICE HISTORY RECORDS (DHR'S) AND PRODUCT COMPLAINT REPORT (PCR) DATABASE RECORDS SHOW THAT THE REPORTED COMPONENT USED IN THE PREVIOUS SURGERY MET DESIGN AND MANUFACTURING REQUIREMENTS. THERE WERE NO NON-CONFORMING MATERIAL REPORTS (NCMR'S) ASSOCIATED WITH THE PRODUCT THAT MAY HAVE CONTRIBUTED TO AN INFECTION. THE DEVICE WAS VERIFIED TO HAVE GONE THROUGH AN ACCEPTABLE STERILIZATION PROCESS AND WAS WITHIN IT'S EXPIRATION DATE AT THE TIME OF USE DURING THE PREVIOUS SURGERY. THERE WERE NO FINDINGS DURING THIS INVESTIGATION THAT INDICATE THAT THE REPORTED DEVICES WERE THE SOURCE/ROOT CAUSE OR HAD A DIRECT CONNECTION WITH THE PATIENT'S INFECTION. THERE ARE MULTIPLE FACTORS THAT MAY CONTRIBUTE TO AN INFECTION THAT ARE OUTSIDE OF THE CONTROL OF DJO SURGICAL. THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUE AFFECTING IMPLANT SAFETY OR EFFECTIVENESS.

Description of Event or Problem · 1

REVISION SURGERY - DUE TO THE PATIENT HAVING A CHRONIC SHOULDER INFECTION. THE SURGEON NOTICED THE INFECTION DURING A PRIOR SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
378318 RSP SHOULDER RSP GLENOID HEAD W/RETAINING SCREW / -4MM OFFSET 32MM KWS ENCORE MEDICAL, L.P. 864C1168

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other| R 506-03-126,LOT 834C1058| 806-03-118,LOT 832C1036| 508-00-008,LOT 854C1024| 506-03-122,LOT 833C1053