ACS POLY W/ 10 LIP 32 GR2
Report
- Report Number
- 1818910-2014-22258
- Event Type
- Injury
- Date Received
- June 27, 2014
- Date of Event
- April 27, 2009
- Report Date
- June 24, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.1818910
- Product Code
- JDI
- PMA / PMN Number
- PK861979
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE LOT CODES REQUIRED WERE NOT PROVIDED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT. THE INVESTIGATION HAS DETERMINED THE LINER PRODUCT CODE IS NOW OBSOLETE. IT WOULD NOT BE UNREASONABLE TO FIND POLY MATERIAL WEAR AFTER THE LENGTH OF TIME REPORTED AS IMPLANTED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
CLINICAL REPORT STATES THAT PATIENT WAS REVISED TO ADDRESS POLY WEAR AND DISLOCATION 22.2 YEARS AFTER PRIMARY THA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 377446 | ACS POLY W/ 10 LIP 32 GR2 | HIP ACETABULAR INSERT/LINER | JDI | DEPUY ORTHOPAEDICS, INC.1818910 | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |