FDA Adverse Event Malfunction Summary report: N

FOLFUSOR

MDR report key: 3900670 · Received June 27, 2014

Report

Report Number
1416980-2014-20736
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
April 23, 2014
Report Date
June 4, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MANUFACTURED SEPTEMBER 5, 2013 ¿ SEPTEMBER 6, 2013. UPON FOLLOW UP WITH THE CUSTOMER IT WAS FOUND THAT THE INFUSION WAS EXPECTED TO HAVE OCCURRED OVER 48 HOURS, HOWEVER, IT WAS COMPLETED IN APPROXIMATELY 31 HOURS. EVALUATION SUMMARY: THE SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION ON THE UNIT SHOWED NO SIGNS OF PHYSICAL ABNORMALITY THAT COULD HAVE CAUSED THE REPORTED CONDITION. A FUNCTIONAL FLOW RATE TEST WAS PERFORMED ON THE UNIT. THE FLOW RATES WERE FOUND TO BE WITHIN THE SPECIFICATION RANGE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE EXACT OCCURRENCE DATE IS UNKNOWN; HOWEVER, THIS WAS REPORTED TO HAVE OCCURRED (B)(6) 2014. A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SMALL VOLUME FOLFUSOR OVERINFUSED. THE REPORTER STATED THAT THE EXPECTED THERAPY TIME WAS 48 HOURS; HOWEVER, THE DEVICE COMPLETED INFUSION IN LESS THAN 48 HOURS. THE DEVICE HAD BEEN FILLED WITH AN UNSPECIFIED DRUG IN 120ML OF 0.9% SODIUM CHLORIDE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 12 OF 19.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
378020 FOLFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 13J010

Patients

Seq Age Sex Outcome Treatment
1 UNSPECIFIED DRUG IN 120ML OF 0.9% SODIUM CHLORIDE