FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3900448 · Received June 27, 2014

Report

Report Number
2955842-2014-03955
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
May 20, 2014
Report Date
June 3, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINED THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. NO OTHER DAMAGE WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY SURGICAL PROCEDURE, IT WAS NOTED THAT A BROKEN CABLE WAS OBSERVED ON THE PROGRASP FORCEPS INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
377129 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-10 M10131025 421

Patients

Seq Age Sex Outcome Treatment
1 54 YR