ENDURANT II
Report
- Report Number
- 2953200-2014-01300
- Event Type
- Injury
- Date Received
- June 27, 2014
- Date of Event
- June 1, 2014
- Report Date
- June 5, 2014
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(40. EXACT EVENT DATE UNK.
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 5CM ABDOMINAL AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS THE DIAMETER OF THE PROXIMAL NECK IS 26MM AND 20MM IN LENGTH. DEGREE OF ANGULATION WAS NOTED AS 10 DEGREES. VESSEL CALCIFICATION WAS NOTED AS MILD. IT WAS REPORTED THAT THE PATIENT PRESENTED WITH LEG PAIN AND CLAUDICATION. THE PHYSICIAN PERFORMED AN ULTRASOUND WHICH IDENTIFIED AN OCCLUSION AT THE LEFT LIMB AND EXTENDED ALL THE WAY TO THE ENTIRE CENTRAL LATERAL SIDE. THE PHYSICIAN TREATED THE PATIENT WITH THROMBOLYSIS FOR 24HRS AND PERFORMED AN INTERVENTION WHERE ANOTHER MANUFACTURER'S BARE METAL STENT WAS IMPLANTED. THE PHYSICIAN NOTED THE CAUSE OF THE OCCLUSION DUE TO THE STENT GRAFT BEING COMPRESSED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS DOING FINE. A REVIEW OF RETURNED ANGIO IMAGES 2-MONTHS POST-IMPLANT REVEALED THE BIFURCATE WAS POSITIONED JUST BELOW THE RENALS. THE IPSI LIMB WAS PLACED INTO THE RIGHT COMMON ILIAC, AND THE CONTRA LIMB INTO THE LEFT COMMON ILIAC ARTERY; THE LIMBS WERE CROSSED. THE CONTRA LIMB WAS COMPLETELY OCCLUDED BEGINNING AT THE FLOW DIVIDER. THE CAUSE OF THE OCCLUSION COULD NOT BE IDENTIFIED FROM THESE IMAGES. FILMS FROM THE SECONDARY PROCEDURE SHOWED THAT THE OCCLUSION WAS RESOLVED FOLLOWING THROMBOLYSIS AND PLACEMENT OF A STENT INSIDE THE CONTRA/LEFT LIMB. THE CAUSE OF THE LIMB OCCLUSION COULD NOT BE CONFIRMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 376570 | ENDURANT II | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V04170477 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention |