FDA Adverse Event Malfunction Summary report: N

ENDO ILS, 29MM, CURVED

MDR report key: 3900224 · Received June 27, 2014

Report

Report Number
3005075853-2014-04479
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 3, 2014
Report Date
June 9, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K940967
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS FOUND THAT THE ECS29A DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE ANVIL HALF WASHER WAS PRESENT AND CUT INDICATING THAT THE INSTRUMENT ACHIEVED A FULL FIRING STROKE; HOWEVER THE INSTRUMENT WAS RECEIVED FULLY LOADED WITH STAPLES. A NEW WASHER WAS PLACED ON THE INSTRUMENT AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE ANVIL WAS ATTACHED ON THE DEVICE COMPLETELY AND WITHOUT ANY DIFFICULTIES AND DID NOT DETACH DURING FUNCTIONAL TESTING. WHILE NO CONCLUSION COULD BE REACHED AS TO HOW THE WASHER WAS CUT WITHOUT THE INSTRUMENT DEPLOYING THE STAPLES, IT IS POSSIBLE THAT THE RETURNED ANVIL DOES NOT BELONG TO THE RETURNED DEVICE. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, THE SURGEON ATTEMPTED TO CONNECT THE ANVIL TO THE STAPLER AND THE TWO COMPONENTS WOULD NOT LOCK TOGETHER. THE SURGEON REMOVED THE STAPLER AND OPENED A NEW DEVICE OF SAME CODE AND WAS ABLE TO CONNECT THE NEW STAPLER TO THE FIRST ANVIL THAT WAS ALREADY IN PLACE IN THE PATIENT. THE CASE WAS COMPLETED WITH FIRST ANVIL AND NEW STAPLER. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
378132 ENDO ILS, 29MM, CURVED STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4EG2Z