EON
Report
- Report Number
- 1627487-2014-21339
- Event Type
- Injury
- Date Received
- May 28, 2014
- Date of Event
- May 6, 2014
- Report Date
- May 6, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AN IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2014-21340. THE PATIENT REPORTED EXPERIENCING UNCOMFORTABLE POCKET HEATING WHILE CHARGING HIS IPG. THE PATIENT HAS ANOTHER MANUFACTURER'S STIMULATOR ON THE OPPOSITE SIDE OF HIS SCS SYSTEM AND FEELS HIS IPG GETS WARM WHILE CHARGING HIS OTHER STIMULATOR AS WELL. A REPLACEMENT CHARGING SYSTEM WAS SENT TO THE PATIENT. ON 08/01/2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314995 | EON | SCS IPG | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3716 | 3083982 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Other | SCS LEAD: MODEL 3228| IMPLANT DATE: |