RELOAD-CURVED STAPLER
Report
- Report Number
- 3005075853-2014-04476
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- May 29, 2014
- Report Date
- June 3, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K040038
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULTS: INCOMPLETE FIRING CYCLE. THE ANALYSIS RESULTS SHOWED THAT THE CR40B RELOAD WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH ONLY 15 STAPLES PRESENT, WITH THE WASHER UNCUT AND WITH THE KNIFE RECESS BELOW THE RELOAD DECK. THE STAPLES WERE NOTED TO BE PROTRUDING; THIS CONDITION IS CONSISTENT WITH THE DEVICE BEING PARTIALLY ACTIVATED. DO NOT FIRE THE INSTRUMENT UNLESS THE CLOSURE TRIGGER IS PROPERLY LATCHED AGAINST THE HANDLE. THE FIRING TRIGGER MUST BE PULLED BACK COMPLETELY AGAINST THE CLOSURE TRIGGER TO PROPERLY FIRE THE INSTRUMENT. IN ADDITION, IF THE FIRING SEQUENCE IS NOT COMPLETE, YOU COULD DEPLOY THE STAPLES WITHOUT CUTTING THE WASHER AND FORMING THE STAPLES. PLEASE REFERENCE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. NO FUNCTIONAL TEST WAS PERFORMED DUE TO THE CONDITION OF THE RELOAD. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAP LOW ANTERIOR RESECTION, AN INSTRUMENT LOADING THE REPORTED CR40B COULD NOT BE FIRED AT THE 2ND FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 377960 | RELOAD-CURVED STAPLER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4AJ8Z |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |