FDA Adverse Event Malfunction Summary report: N

TRANSEND EX 014/205 PLATINUM

MDR report key: 3899680 · Received June 27, 2014

Report

Report Number
3008853977-2014-00198
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
June 5, 2014
Report Date
June 10, 2014
Manufacturer
BOSTON SCIENTIFIC COSTA RICA
Product Code
DQX
PMA / PMN Number
K971254
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE GUIDEWIRE WAS RECEIVED IN TWO PIECES: THE PROXIMAL SECTION WAS MEASURED 94.5CM AND THE DISTAL SECTION WAS MEASURED 108.5CM. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT BOTH PIECES WERE BENT ON SEVERAL PLACES ALONG ITS LENGTH. MAGNIFIED EXAMINATION OF THE FRACTURE SITE SHOWED THE CORE WIRE WAS FRACTURED. NO BENDING AND OR STRETCHING ON THE GUIDEWIRE WAS NOTICED. FROM THE CONDITION OF THE GUIDEWIRE AT THE FRACTURE SITE, IT APPEARS THAT THE GUIDEWIRE WAS SEPARATED DUE TO OVER TORQUE. THE BENDING ON THE GUIDEWIRE APPEARED TO BE DUE TO EXCESSIVE MANIPULATION. THE FRACTURE ON THE GUIDEWIRE APPEARED TO BE DUE TO THE EXCESSIVE MANIPULATION AND TORSIONAL OVERLOAD. HOWEVER, REVIEW AND ANALYSIS OF AVAILABLE INFORMATION AND INVESTIGATION RESULTS FAILED TO IDENTIFY A DEFINITIVE ROOT CAUSE FOR THE REPORTED EVENT AND OBSERVED ISSUES. THE GUIDEWIRE POLYTETRAFLUOROETHYLENE (PTFE) COATING WAS FOUND TO BE SCRAPED OFF ON SEVERAL PLACES ALONG ITS PROXIMAL APPROXIMATELY 38.0CM LENGTH. THE PTFE COATING APPEARED TO BE SCRAPED OFF OF THE GUIDEWIRE WITH THE TORQUE DEVICE COLLET. NO ANOMALIES WERE OBSERVED WITH THE HYDROPHILIC COATING. THE DIRECTIONS FOR USE (DFU) CAUTIONS TO: "EXCESSIVE TIGHTENING OF THE TORQUE DEVICE ONTO THE WIRE MAY RESULT IN ABRASION OF THE COATING OF THE WIRE." THE COATING DAMAGE ON THE PROXIMAL END OF THE GUIDEWIRE WAS PROBABLY DUE TO THE INSUFFICIENT TIGHTENING OF THE TORQUE DEVICE. THEREFORE, A ROOT CAUSE OF DFU INSTRUCTIONS NOT FOLLOWED HAS BEEN ASSIGNED TO THIS OBSERVATION.

Additional Manufacturer Narrative · 1

THE DEVICE IS NOT AVAILABLE TO THE MANUFACTURER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TORQUING THE GUIDEWIRE APPROXIMATELY 3-5 CM OF THE DISTAL TIP BROKE OFF. THE GUIDEWIRE WAS COMPLETELY REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT DUE TO THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TORQUING THE GUIDEWIRE APPROXIMATELY 3-5 CM OF THE DISTAL TIP BROKE OFF. THE GUIDEWIRE WAS COMPLETELY REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT DUE TO THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376724 TRANSEND EX 014/205 PLATINUM WIRE, GUIDE, CATHETER DQX BOSTON SCIENTIFIC COSTA RICA 16469331

Patients

Seq Age Sex Outcome Treatment
1 TRANSFORM OCCLUSION BALLOON CATHETER (STRYKER)