FDA Adverse Event Malfunction Summary report: N

PACKER/CHANG IOL CUTTER

MDR report key: 3899031 · Received May 20, 2014

Report

Report Number
3019924-2014-00017
Event Type
Malfunction
Date Received
May 20, 2014
Report Date
May 20, 2014
Manufacturer
MICROSURGICAL TECHNOLOGY, INC.
Product Code
HNF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE SCISSOR WAS RETURNED ALONG WITH THE ONE PIECE HARD IOL (UNCUT) WITH DAMAGE THAT IS INDICATIVE OF THE SCISSOR BEING USED TO CUT TOO HARD OF AN OBJECT. THE ONE PIECE LENS WAS SCORED WHERE THE SURGEON ATTEMPTED TO CUT THE LENS. THE SURGEON REPORTED THAT HE WAS ATTEMPTING TO USE THE SCISSOR TO CUT THE ONE PIECE ACRYLIC IOL. THIS SCISSOR IS INDICATED TO CUT SOFT FOLDABLE IOLS.

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED THAT THE PACKER/CHANG IOL SCISSOR BROKE WHEN THE SURGEON ATTEMPTED TO CUT A ONE PIECE ACRYLIC IOL. THERE WAS NO IMPACT TO THE PATIENT AND THE BROKEN PIECE WAS REMOVED FROM THE EYE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299556 PACKER/CHANG IOL CUTTER OPHTHALMIC SCISSOR HNF MICROSURGICAL TECHNOLOGY, INC. DFH-0012 050336

Patients

Seq Age Sex Outcome Treatment
1 80 YR