FDA Adverse Event
Malfunction
Summary report: N
PACKER/CHANG IOL CUTTER
MDR report key: 3899031
·
Received May 20, 2014
Report
- Report Number
- 3019924-2014-00017
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Report Date
- May 20, 2014
- Manufacturer
- MICROSURGICAL TECHNOLOGY, INC.
- Product Code
- HNF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE SCISSOR WAS RETURNED ALONG WITH THE ONE PIECE HARD IOL (UNCUT) WITH DAMAGE THAT IS INDICATIVE OF THE SCISSOR BEING USED TO CUT TOO HARD OF AN OBJECT. THE ONE PIECE LENS WAS SCORED WHERE THE SURGEON ATTEMPTED TO CUT THE LENS. THE SURGEON REPORTED THAT HE WAS ATTEMPTING TO USE THE SCISSOR TO CUT THE ONE PIECE ACRYLIC IOL. THIS SCISSOR IS INDICATED TO CUT SOFT FOLDABLE IOLS.
Description of Event or Problem · 1
THE DISTRIBUTOR REPORTED THAT THE PACKER/CHANG IOL SCISSOR BROKE WHEN THE SURGEON ATTEMPTED TO CUT A ONE PIECE ACRYLIC IOL. THERE WAS NO IMPACT TO THE PATIENT AND THE BROKEN PIECE WAS REMOVED FROM THE EYE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 299556 | PACKER/CHANG IOL CUTTER | OPHTHALMIC SCISSOR | HNF | MICROSURGICAL TECHNOLOGY, INC. | DFH-0012 | 050336 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |