FDA Adverse Event
Summary report: N
RELIANCE SYNERGY WASHER
MDR report key: 3897184
·
Received June 26, 2014
Report
- Report Number
- 9680353-2014-00042
- Date Received
- June 26, 2014
- Date of Event
- June 9, 2014
- Report Date
- June 26, 2014
- Manufacturer
- STERIS CANADA CORPORATION
- Product Code
- MEC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A STERIS SERVICE TECHNICIAN INSPECTED THE WASHER AND FOUND THE DRYING PISTON VALVE TO BE BROKEN SUBSEQUENTLY ALLOWING THE HOSE TO DETACH AND WATER TO LEAK. THE TECHNICIAN REPAIRED THE WASHER BY REPLACING THE DRYING PISTON VALVE AND O-RINGS, RAN A TEST CYCLE AND CONFIRMED THE WASHER TO BE OPERATIONAL. NO FURTHER ISSUES HAVE BEEN REPORTED.
Description of Event or Problem · 1
THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR RELIANCE SYNERGY WASHER. NO INJURIES OR PROCEDURAL DELAYS/CANCELLATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 373366 | RELIANCE SYNERGY WASHER | WASHER | MEC | STERIS CANADA CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |