FDA Adverse Event Summary report: N

RELIANCE SYNERGY WASHER

MDR report key: 3897184 · Received June 26, 2014

Report

Report Number
9680353-2014-00042
Date Received
June 26, 2014
Date of Event
June 9, 2014
Report Date
June 26, 2014
Manufacturer
STERIS CANADA CORPORATION
Product Code
MEC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE WASHER AND FOUND THE DRYING PISTON VALVE TO BE BROKEN SUBSEQUENTLY ALLOWING THE HOSE TO DETACH AND WATER TO LEAK. THE TECHNICIAN REPAIRED THE WASHER BY REPLACING THE DRYING PISTON VALVE AND O-RINGS, RAN A TEST CYCLE AND CONFIRMED THE WASHER TO BE OPERATIONAL. NO FURTHER ISSUES HAVE BEEN REPORTED.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM THEIR RELIANCE SYNERGY WASHER. NO INJURIES OR PROCEDURAL DELAYS/CANCELLATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373366 RELIANCE SYNERGY WASHER WASHER MEC STERIS CANADA CORPORATION

Patients

Seq Age Sex Outcome Treatment
1