FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3897156 · Received May 21, 2014

Report

Report Number
8020893-2014-01224
Event Type
Malfunction
Date Received
May 21, 2014
Report Date
April 24, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED TO HAVE REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB). THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) UPLOADED THE OPERATIONAL SOFTWARE ONLY AND PERFORMED THE DEVICE EVALUATION. THE VENTILATOR PASSED THE EXTENDED SELF-TEST (EST), SHORT SELF-TEST (SST) AND THE PERFORMANCE VERIFICATION TEST (PVT); THE DEVICE OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS. COVIDIEN REFERENCE (B)(4).

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION FROM A CUSTOMER IN THE USA STATING THAT AN 840 VENTILATOR HAD AN ERRATIC DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE EXACT DATE OF THE INCIDENT IS UNK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
304007 840 VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND 840

Patients

Seq Age Sex Outcome Treatment
1