FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3897136
·
Received May 21, 2014
Report
- Report Number
- 8020893-2014-01208
- Event Type
- Malfunction
- Date Received
- May 21, 2014
- Date of Event
- April 20, 2014
- Report Date
- April 22, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION ADN REPLACED THE GRAPHICAL USER INTERFACE (GUI) TOUCH FRAME PRINTED CIRCUIT BOARD (PCB). THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS. COVIDIEN REFERENCE: (B)(4).
Description of Event or Problem · 1
A REPORT RECEIVED FROM THE USA STATED AN 940 VENTILATOR EXPERIENCED A BLACK SCREEN. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.