FDA Adverse Event Malfunction Summary report: N

EMERGE?

MDR report key: 3896947 · Received June 26, 2014

Report

Report Number
2134265-2014-03658
Event Type
Malfunction
Date Received
June 26, 2014
Date of Event
June 4, 2014
Report Date
June 4, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
K113220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATED BY MFR.: THE COMPLAINT DEVICE WAS NOT RECEIVED AT THE COMPLAINT INVESTIGATION SITE (CIS) FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED CHRONIC TOTAL OCCLUDED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS UNSPECIFIED VESSEL. A 2.00MM X 15MM EMERGE¿ BALLOON CATHETER WAS ADVANCED TO DILATE THE LESION. ON THE FIRST INFLATION, THE BALLOON RUPTURED AT UNSPECIFIED PRESSURE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS' STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
374144 EMERGE? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493919315200 16756904

Patients

Seq Age Sex Outcome Treatment
1