FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 3896808 · Received May 20, 2014

Report

Report Number
1824206-2014-01556
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BEDS SCALE WAS NOT ZEROED MOST LIKELY DUE TO USER ERROR. A CROOKED LATCH. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012 AND 2013. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN ZEROED THE SCALE TO RESOLVE THE ISSUE. THE TECHNICIAN THOROUGHLY TESTED THE BED EXIT SYSTEM AND IT FUNCTIONED AS DESIGNED. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BED EXIT ALARM WOULD NOT SET. THE BED WAS LOCATED AT THE ACCOUNT IN THE WAREHOUSE. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300632 VERSACARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment