ARTICULEZE M HEAD 36MM +15.5
Report
- Report Number
- 1818910-2014-22082
- Event Type
- Injury
- Date Received
- June 26, 2014
- Date of Event
- May 27, 2014
- Report Date
- May 27, 2014
- Manufacturer
- DEPUY INTERNATIONAL LTD.-8010379
- Product Code
- JDI
- PMA / PMN Number
- PK980513
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DISLOCATED HIP. BROKEN CONSTRAINING RING ON A CONSTRAINED LINER. DOI: (B)(6) 2013, DOR: (B)(6) 2014 (LEFT HIP). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO ADDITIONAL REPORTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THE FK4LY1 LOT CODE DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
DISLOCATED HIP. BROKEN CONSTRAINING RING ON A CONSTRAINED LINER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 373495 | ARTICULEZE M HEAD 36MM +15.5 | HIP FEMORAL HEAD | JDI | DEPUY INTERNATIONAL LTD.-8010379 | 3558011 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention |