FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3896628
·
Received January 16, 2014
Report
- Report Number
- 1314492-2014-03564
- Event Type
- Malfunction
- Date Received
- January 16, 2014
- Date of Event
- December 1, 2013
- Report Date
- December 20, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MFR REF NO: (B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE REPORTED SYMPTOM SYSTEM ERROR 322 WAS CONFIRMED. A PHYSICAL INSPECTION OF THE DEVICE FOUND THAT THE UPPER AND LOWER LATCH SWITCH BRACKET SCREWS WERE LOOSE ALLOWING THE UPPER AND LOWER LATCH SWITCH TO MOVE IN AND OUT FROM THE UPPER AND LOWER DOOR LATCH. THE LOOSE NYLON SCREWS WILL TRIGGER AN INTERMITTENT OPEN/CLOSED SWITCH CONNECTION CAUSING THE SYSTEM ERROR 322. AS A RESULT, THE UPPER AUXILIARY ASSEMBLY WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP WAS ALARMING FOR A SYSTEM ERROR 322. THERE WAS NO PT INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 43598 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |