FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3896615
·
Received May 19, 2014
Report
- Report Number
- 1720753-2014-04288
- Event Type
- Malfunction
- Date Received
- May 19, 2014
- Date of Event
- May 12, 2014
- Report Date
- May 19, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEM (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER HAS REPORTED SEEING CONTROL PANEL ERROR MESSAGE AND SYSTEM WILL NOT SCREEN. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THIS COULD CAUSE THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NOT REPORT OF INJURY OR DEATH ASSOCIATED WITH EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 296284 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEM (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |