MAESTRO DRILL WITH HANDSWITCH
Report
- Report Number
- 0001811755-2014-02278
- Event Type
- Malfunction
- Date Received
- June 26, 2014
- Date of Event
- May 30, 2014
- Report Date
- May 30, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
DURING FAILURE ANALYSIS, THE REPORTED EVENT OF THE DEVICE RUNNING WITHOUT USER ACTIVATION WAS CONFIRMED. THE DEVICE HAD A DAMAGED O-RING ON THE NEEDLE VALVE. DAMAGE TO THE O-RING CAN CAUSE THE NEEDLE VALVE TO STICK, KEEPING THE DEVICE ACTIVATED IN SAFE MODE.
FAILURE ANALYSIS IS IN PROGRESS. A FOLLOW-UP WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETE. H3 OTHER TEXT : DEVICE EVALUATION IN PROGRESS
IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING SERVICE EVALUATION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.
IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING SERVICE EVALUATION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 373460 | MAESTRO DRILL WITH HANDSWITCH | MOTOR, DRILL, PNEUMATIC | HBB | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |