MAESTRO DRILL WITH HANDSWITCH

MDR report key: 3896608 · Received June 26, 2014

Report

Report Number
0001811755-2014-02278
Event Type
Malfunction
Date Received
June 26, 2014
Date of Event
May 30, 2014
Report Date
May 30, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING FAILURE ANALYSIS, THE REPORTED EVENT OF THE DEVICE RUNNING WITHOUT USER ACTIVATION WAS CONFIRMED. THE DEVICE HAD A DAMAGED O-RING ON THE NEEDLE VALVE. DAMAGE TO THE O-RING CAN CAUSE THE NEEDLE VALVE TO STICK, KEEPING THE DEVICE ACTIVATED IN SAFE MODE.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS. A FOLLOW-UP WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETE. H3 OTHER TEXT : DEVICE EVALUATION IN PROGRESS

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING SERVICE EVALUATION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH WAS RUNNING WITHOUT USER ACTIVATION DURING SERVICE EVALUATION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.