SPECTRUM INFUSION PUMP
MDR report key: 3896559
·
Received January 16, 2014
Report
- Report Number
- 1314492-2014-03550
- Event Type
- Malfunction
- Date Received
- January 16, 2014
- Date of Event
- December 1, 2013
- Report Date
- December 20, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MFR REF NO: (B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND TO BE OUT OF SPEC IN RELATION TO THE REPORTED SYMPTOMS, WHICH WAS REPRODUCED. SYSTEM ERROR 322 ALARM WERE CONFIRMED AND WERE DETERMINED TO BE CAUSED BY A LOOSE UPPER LATCH SWITCH BRACKET. THE UPPER AUXILIARY ASSEMBLY, WHICH CONTAINS THE LOOSE UPPER LATCH SWITCH BRACKET, WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP ALARMED SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.