FDA Adverse Event Malfunction Summary report: N

NEXGEN MIS STEMMED TIBIAL COMPONENT

MDR report key: 3896528 · Received May 19, 2014

Report

Report Number
2648920-2014-00128
Event Type
Malfunction
Date Received
May 19, 2014
Date of Event
September 13, 2013
Report Date
September 13, 2013
Manufacturer
ZIMMER
Product Code
JWH
Removal / Correction Number
2648920-5/19-2014-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PLEASE REFERENCE MDR #1822565-2013-01469 FOR THE INITIAL REPORT SUBMITTED FOR THIS EVENT. NO OUT OF SPECIFICATION CONDITION HAS BEEN CONFIRMED FOR THE MIS TIBIAL COMPONENT REPORTED IN THIS EVENT. HOWEVER, REVIEW OF PRODUCT FOR MDR #2648920-2013-00272 FOUND THAT REPORTED LOT TO HAVE MATING THREADS WHICH ARE OUT OF SPECIFICATION. THE SCOPE OF THE INVESTIGATION WAS EXPANDED TO INCLUDE THE TIBIAL COMPONENT REPORTED FOR THIS EVENT. AS A RESULT, FIELD ACTION 2648920-05/19/2014-001-R IS BEING INITIATED. THIS REMOVAL REPORT CAN BE REFERENCED FOR ADDITIONAL INFORMATION.

Description of Event or Problem · 1

IT IS REPORTED THAT THE SURGEON WAS UNABLE TO INSERT THE DROP DOWN STEM INTO THE MIS TIBIAL COMPONENT, SO THE SURGERY WAS COMPLETED WITH ANOTHER DROP DOWN STEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
297698 NEXGEN MIS STEMMED TIBIAL COMPONENT JWH ZIMMER 62261962

Patients

Seq Age Sex Outcome Treatment
1