FDA Adverse Event Malfunction Summary report: N

TERUMP TCM II COOLING AND HEATING SYSTEM

MDR report key: 3896490 · Received May 16, 2014

Report

Report Number
1828100-2014-00443
Event Type
Malfunction
Date Received
May 16, 2014
Date of Event
April 1, 2014
Report Date
April 24, 2014
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS DETERMINED BY TECHNICAL SUPPORT TO BE A FAULTY PRIME PSI SWITCH. TECHNICAL SUPPORT GAVE THE CUSTOMER, IN AN EMAIL, THE PART NUMBER OF THE PRIME PSI SWITCH AND THE PRICE, ALONG WITH THE INSTRUCTIONS ON HOW TO CHANGE THE SWITCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT DEVELOPED A PRIME FAULT. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, OR ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
294603 TERUMP TCM II COOLING AND HEATING SYSTEM TCM II COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4415