Narratives
Additional Manufacturer Narrative · 1
IT WAS DETERMINED BY TECHNICAL SUPPORT TO BE A FAULTY PRIME PSI SWITCH. TECHNICAL SUPPORT GAVE THE CUSTOMER, IN AN EMAIL, THE PART NUMBER OF THE PRIME PSI SWITCH AND THE PRICE, ALONG WITH THE INSTRUCTIONS ON HOW TO CHANGE THE SWITCH.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT DEVELOPED A PRIME FAULT. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, OR ADVERSE CONSEQUENCES TO THE PATIENT.