UNKNOWN DEPUY ACETABULAR CUP
Report
- Report Number
- 1818910-2014-22077
- Event Type
- Injury
- Date Received
- June 26, 2014
- Date of Event
- May 27, 2014
- Report Date
- May 27, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- KWA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
PATIENT WAS REVISED TO ADDRESS DISASSOCIATION. VISUAL EXAMINATION OF THE RETURNED DEVICES FINDS EVIDENCE TO SUPPORT DISASSOCIATION OF THE LINER AND CUP WHILE IMPLANTED. THE RETURNED FEMORAL HEAD EXHIBITS SIGNS OF COMING INTO DIRECT CONTACT WITH THE ACETABULAR CUP, FURTHER SUPPORTING THE DISASSOCIATION. IN THIS CASE 4 OF THE ARD'S OF THE POLYETHYLENE LINER HAVE FRACTURED. IT IS EVIDENT THAT THE EDGE OF THE LINER WAS LOADED BY THE FEMORAL HEAD AND PATIENT BODY WEIGHT WHILE IMPLANTED. THERE ARE MACHINING LINES YET VISIBLE WITHIN THE INNER RADIUS OF THE LINER WHICH WOULD NOT BE AS OBVIOUSLY PRESENT HAD THE FEMORAL HEAD BEEN MORE CENTRALLY LOADED. THE PATIENT IS A REPORTED MALE, (B)(6). THE PATIENT IS CONSIDERED OBESE. IT IS STATED IN THE WARNINGS AND PRECAUTIONS THAT EXCESSIVE PATIENT WEIGHT TENDS TO ADVERSELY AFFECT HIP REPLACEMENT IMPLANTS. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE PRODUCT AND LOT CODE COMBINATION SINCE ITS RELEASE TO DISTRIBUTION. ADDITIONAL INFORMATION AND INVESTIGATIONAL INPUTS INCLUDING X-RAYS WERE REQUESTED BUT NOT PROVIDED. PLACEMENT CANNOT BE COMMENTED ON. THE INVESTIGATION CAN DRAW NO FURTHER CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THIS INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
PATIENT WAS REVISED TO ADDRESS DISASSOCIATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 374925 | UNKNOWN DEPUY ACETABULAR CUP | HIP ACETABULAR CUP | KWA | DEPUY ORTHOPAEDICS, INC. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention |