FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3896413 · Received January 16, 2014

Report

Report Number
1314492-2014-03498
Event Type
Malfunction
Date Received
January 16, 2014
Date of Event
December 1, 2013
Report Date
December 20, 2013
Manufacturer
BAXTER HEALTHCARE CORP.
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO.:(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE "DOWNSTREAM OCCLUSION" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFO IS RECEIVED A FOLLOW UP WILL BE SENT. THE FORCE SENSOR PUSHER, FORCE SENSOR GASKET AND DOWNSTREAM TUBING GUIDE WERE REPLACED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 199 OCCURRENCES OF "DOWNSTREAM OCCLUSION" ALARM WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE. NO ADDITIONAL INFO IS AVAILABLE.