FDA Adverse Event Malfunction Summary report: N

ADVANCE BED

MDR report key: 3896359 · Received May 20, 2014

Report

Report Number
1824206-2014-01585
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 28, 2014
Report Date
April 28, 2014
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE RIGHT HEAD BRAKE CASTER NOT HOLDING DUE TO IT NEEDING TO BE ADJUSTED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2008-2013. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN ADJUSTED THE BRAKE CASTER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE RIGHT HEAD BRAKE CASTER IS NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN BED REPAIR. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300829 ADVANCE BED AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1135

Patients

Seq Age Sex Outcome Treatment
1