FDA Adverse Event Malfunction Summary report: N

SELENIA DIMENSION MAMMOGRAPHY SYSTEM 3D

MDR report key: 3896353 · Received May 20, 2014

Report

Report Number
1220984-2014-00004
Event Type
Malfunction
Date Received
May 20, 2014
Report Date
May 20, 2014
Manufacturer
HOLOGIC INC
Product Code
OTE
PMA / PMN Number
P080003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE HOLOGIC FIELD ENGINEER WENT TO THE SITE ON (B)(4) 2014 AND FOUND THEM USING A ROLLING X-RAY SHIELD. THE BROKEN PIECES OF GLASS THAT HAD COME FROM THE SELENIA DIMENSIONS SYSTEM HAD ALREADY BEEN DISCARDED, SO THERE WAS NOTHING AVAILABLE TO EXAMINE. THERE WAS NO REPORT OF ANYONE BUMPING OR OTHERWISE DAMAGING THE SHIELD. THE FIELD ENGINEER REPLACED THE X-RAY SHIELD. THE X-RAY SHIELD PROVIDED WITH THE SELENIA DIMENSIONS SYSTEM IS MADE OF TEMPERED GLASS (SAFETY GLASS), SO AS NOT TO CAUSE HARM IF THE GLASS IS BROKEN. THIS PARTICULAR LOT OF GLASS WAS PREVIOUSLY EVALUATED AND DETERMINED TO MEET THE REQUIRED SPECIFICATIONS.

Description of Event or Problem · 1

THE TECHNOLOGIST LEFT FOR THE DAY AND WHEN THEY RETURNED FOUND THE X-RAY SHIELD ON THE SELENIA DIMENSIONS SYSTEM HAD SHATTERED. IT HAD BEEN CLEANED UP BY THE TIME THE TECHNOLOGIST RETURNED, SO THEY COULD NOT DESCRIBE THE CIRCUMSTANCES REGARDING WHAT HAPPENED.