FDA Adverse Event Malfunction Summary report: N

ADVANCE BED

MDR report key: 3896323 · Received May 20, 2014

Report

Report Number
1824206-2014-01567
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CASTERS ARE WORN DUE TO NORMAL WEAR AND TEAR. PER THE HILL-ROM SERVICE MANUAL THE ADVANCE SERIES BED REQUIRES EFFECTIVE MAINTENANCE PROGRAM. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. CHECK THE TIRES FOR CUTS, WEAR, TREAD LIFE, ETC. TEST THE BRAKE CASTERS TO DETERMINE IF THE BED MOVES WHEN YOU ACTIVATE THE BRAKE MODE, REPLACE OR ADJUST WHEN NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2006 THROUGH 2008. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE BRAKE CASTERS AND BRAKE STEER CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN REPORTED THE BRAKES WILL SET BUT DO NOT HOLD. THE BED WAS LOCATED IN STORAGE AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300963 ADVANCE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1105

Patients

Seq Age Sex Outcome Treatment
1