AFFINITY THREE BED
Report
- Report Number
- 1824206-2014-01543
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- April 22, 2014
- Report Date
- April 22, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- PMA / PMN Number
- K915779
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
THE ACCOUNT FOUND THE BRAKE DETENT WAS BROKEN. PER THE HILL-ROM SERVICE MANUAL THE AFFINITY THREE BIRTHING BED AND AFFINITY FOUR BIRTHING BEDS REQUIRE AND EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM SEMIANNUAL PREVENTIVE MAINTENANCE. CHECK THE TIRES FOR CUTS, WEAR, TREAD LIFE, ETC. APPLY THE BRAKE, AND CHECK TO ENSURE THAT THE BED WILL NOT MOVE. IF THE BED MOVES, INSPECT IT FOR WEAR, AND ADJUST IF REQUIRED. APPLY THE STEERING PEDAL AND CHECK THE STEERING TO ENSURE PROPER LOCKING ACTION WHEN ACTIVATED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2010-2012. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE ACCOUNT REPLACED THE BRAKE DETENT TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILE DIN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300854 | AFFINITY THREE BED | BIRTHING BED | HDD | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |