FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 3896294 · Received May 20, 2014

Report

Report Number
1824206-2014-01550
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K962942
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT DID NOT PERFORM A FAILURE ANALYSIS BUT DID FIND THE CAUSE TO BE THE POWER CONTROL MODULE POWER CONTROL BOARD ASSEMBLY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON ITS BEDS. THE ACCOUNT REPLACED THE POWER CONTROL MODULE POWER CONTROL BOARD ASSEMBLY TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THAT THE BED EXIT ALARM WILL SOUND AT THE NURSE'S STATION BUT NOT AT THE BED. THE BED WAS LOCATED IN THE REPAIR AREA AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300715 TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1