FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3896261 · Received January 16, 2014

Report

Report Number
1314492-2014-03434
Event Type
Malfunction
Date Received
January 16, 2014
Date of Event
December 1, 2013
Report Date
December 20, 2013
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO: (B)(4). THE DEVICE WAS RETURNED TO AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE UPSTREAM OCCLUSION ALARMS WERE CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP WILL BE SENT. THE PROCESSOR PRINTED CIRCUIT BOARD WAS REPLACED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 9 OCCURRENCES OF "UPSTREAM OCCLUSION" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
43896 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1