FDA Adverse Event Summary report: N

SPINE & TRAUMA ICT; SW VERSION 1.0.2

MDR report key: 3896167 · Received June 12, 2014

Report

Report Number
8043933-2014-00020
Date Received
June 12, 2014
Date of Event
May 16, 2014
Report Date
May 20, 2014
Manufacturer
BRAINLAB AG
Product Code
OLO
PMA / PMN Number
K083310
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A RISK TO THE PATIENT'S HEALTH COULD NOT BE EXCLUDED FOR THESE SPECIFIC CIRCUMSTANCES, ALTHOUGH THERE IS NO INDICATION OF A MALFUNCTION OR DEFECT OR (SYSTEMATIC) ERROR OF THE BRAINLAB DEVICE, THE BRAINLAB DEVICE WORKS AS INTENDED. THE CORRESPONDING BRAINLAB MEASURES TO MINIMIZE THIS ALREADY ANTICIPATED RISK AS LOW AS REASONABLY PRACTICABLE ARE IN PLACE. ACCORDING TO THE HOSPITAL (SURGEON), THERE IS NO NEGATIVE EFFECT TO THE PATIENT, AND THE FINAL OUTCOME OF THE SURGERY WAS SUCCESSFUL. THE VIRTUAL DISPLAY OF THE BRAINLAB NAVIGATION MIGHT NOT HAVE BEEN AS ACCURATE TO THE PATIENT ANATOMY AS DESIRED DURING THE PLACEMENT OF THESE K-WIRES FOR THE RESULTING PEDICLE SCREWS, CAUSED BY POTENTIAL MOVEMENTS OF THE REFERENCE ARRAY DUE TO NOT SUFFICIENTLY RIGID FIXATION TO THE SPINOUS PROCESS (USER FIXATED IN BETWEEN TWO SPINOUS PROCESSES). OF THE VERTEBRA THAT COULD NOT BE COMPENSATED BY THE NAVIGATION SYSTEM, SINCE THE REFERENCE ARRAY WAS NOT ATTACHED ON THE BONE OPERATED ON. BRAINLAB INVESTIGATION COULD NEITHER CONFIRM NOR DISPROVE AN ACTUAL DISCREPANCY OF THE INFO VIRTUALLY DISPLAYED BY THE BRAINLAB NAVIGATION AND THE ACTUAL CURRENT ANATOMY AND INSTRUMENT POSITIONS. BRAINLAB INTENDS TO RE-ITERATE TO THIS CUSTOMER. THE REQUIREMENT FOR RIGID FIXATION OF THE REFERENCE ARRAY TO THE BONE. THAT ONLY IF THE REFERENCE ARRAY IS ATTACHED ON THE BONE TO BE OPERATED ON (AS REQUIRED IN THE BRAINLAB INSTRUCTIONS FOR USE), POTENTIAL MOVEMENTS OF THE BONE OPERATED ON WOULD NOT LEAD TO DISCREPANCIES BETWEEN VIRTUAL NAVIGATION DISPLAY AND ACTUAL CURRENT OF PATIENT ANATOMY OF INTEREST.

Description of Event or Problem · 1

A MINIMALLY-INVASIVE SPINAL SURGERY FOR A STABILIZATION OF THE THORACIC SPINE (SPONDYLISTHESIS, T3-T7, EXCLUDING T5) HAS BEEN PERFORMED, WITH THE AID OF THE VIRTUAL DISPLAY BY THE BRAINLAB NAVIGATION FOR THE PLACEMENT OF THE K-WIRES FOR RESULTING PEDICLE SCREWS. INTRA-OPERATIVE CT VERIFICATION DURING THIS SURGERY REVEALED THAT 6 OUT OF 8 PLACED SCREWS WERE TILTED BY APPROX 10-15 DEGREES FROM THE INTENDED POSITIONS. THE POSITIONS OF THESE PEDICLE SCREWS WERE CORRECTED DURING THE SAME SURGERY. ACCORDING TO THE HOSPITAL (SURGEON), THERE IS NO NEGATIVE EFFECT TO THE PT, AND THE FINAL OUTCOME OF THE SURGERY WAS SUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
348265 SPINE & TRAUMA ICT; SW VERSION 1.0.2 IMAGE GUIDED SURGERY SYSTEM/STEREOT. OLO BRAINLAB AG 22180B SW VERSION 1.0.2

Patients

Seq Age Sex Outcome Treatment
1 Other