FDA Adverse Event Malfunction Summary report: N

SYMBIQ DCHANNEL 3.01

MDR report key: 3896024 · Received May 15, 2014

Report

Report Number
9615050-2014-03446
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
January 1, 2014
Report Date
April 18, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0072-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE TOUCHSCREEN DID NOT PASS THE TOUCHSCREEN TEST AND DID NOT RESPOND WHEN PRESSED. THE PROBABLE CAUSE WAS DUE TO CONTAMINATION ON THE DEVICE TOUCHSCREEN DUE TO FLUID INGRESS. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WAS NOT WORKING. THE DEVICE WAS RETUREND TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290830 SYMBIQ DCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK