FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SINGLE CHANNE
MDR report key: 3896005
·
Received May 15, 2014
Report
- Report Number
- 9615050-2014-03438
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 10, 2014
- Report Date
- April 16, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0069-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION AT THE SERVICE CENTER FOUND THE DEVICE TOUCHSCREEN DID RESPOND WHEN PRESSED. THE PROBABLE CAUSE WAS DUE TO CONTAMINATION DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DOES NOT CALIBRATE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.