FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3895921 · Received June 26, 2014

Report

Report Number
2939301-2014-15538
Event Type
Malfunction
Date Received
June 26, 2014
Report Date
June 16, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2014, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(4) ALLEGING THE ONETOUCH VERIO IQ METER READ INACCURATELY COMPARED TO ANOTHER DEVICE. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF ¿188, 230, 287, 366 AND 328 MG/DL¿ WITH THE SUBJECT METER AND ¿56, 65, 111, 180 AND 160 MG/DL¿ ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER, RESPECTIVELY. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR METER TO METER ACCURACY TESTING, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
374678 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3579419

Patients

Seq Age Sex Outcome Treatment
1