ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-18344
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Report Date
- June 21, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 11/16/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/28/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE PUMP POWERED UP WITH AUDITORY AND VIBRATORY FEATURES. THE ¿UP¿, ¿DOWN¿ AND ¿OK¿ BUTTONS RESPONDED INTERMITTENTLY. REMOVAL OF THE KEYPAD COVER FOUND CONTAMINATION UNDER ALL THE BUTTON CONTACTS. THE REPORTED BUTTON TACTILE ISSUE WAS DUPLICATED DURING THE INVESTIGATION. UNRELATED TO THE BUTTON ISSUE, THE DISPLAY WAS DIM AND DISCOLORED.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE ¿UP¿, ¿DOWN¿ AND ¿OK¿ BUTTONS WERE UNDER RESPONSIVE. IT WAS REPORTED THAT THE KEYPAD COVER WAS UNDAMAGED AND THERE WAS NO EVIDENCE OF MOISTURE OR CORROSION IN THE PUMP. THE REPORTER DID NOT NOTE ANY DELIVERY ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 372159 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |