ANIMAS VIBE
Report
- Report Number
- 2531779-2014-18320
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Report Date
- June 17, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1: DATE OF SUBMISSION 08/18/2014, DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE BLACK BOX SHOWED SEVERAL LOSS OF PRIME EVENTS DUE TO LOW NON ZERO FORCE CONDITIONS. NO LOSS OF PRIME EVENT WAS DUPLICATED DURING THE INVESTIGATION. THE PRODUCT PERFORMED WITHIN SPECIFICATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THE REPORTER STATED THAT A RECURRENT LOSS OF PRIME WARNING HAD OCCURRED THAT WAS UNRESOLVED WITH A BATTERY CHANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.