FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3895681 · Received June 25, 2014

Report

Report Number
2531779-2014-18293
Event Type
Malfunction
Date Received
June 25, 2014
Report Date
June 23, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/14/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT A ¿NO CARTRIDGE DETECTED¿ ALARM WAS RECORDED. DURING TESTING, THE PUMP POWERED UP AND IMMEDIATELY EMITTED A REPLACE BATTERY ALARM. THE PUMP CASE WAS REMOVED AND CORROSION WAS FOUND ON MULTIPLE COMPONENTS OF THE PRINTED CIRCUIT BOARD AS WELL AS ON THE FORCE SENSOR FLEX CONNECTOR. THE COMPLAINT THAT THE PUMP WAS PRIMING THE TUBING WHEN PERFORMING THE LOAD STEP WAS NOT ABLE TO BE ADEQUATELY INVESTIGATED DUE TO THE REPLACE BATTERY ALARM AND CORROSION INSIDE THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOAD STEP MALFUNCTION) ISSUE. THE REPORTER ALLEGED THAT THE PUMP PRIMED THE TUBING DURING THE LOAD STEP. IT WAS NOTED THAT THE PUMP EMITTED A "NO CARTRIDGE DETECTED" ALARM. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372061 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 10 YR