FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3895664 · Received June 25, 2014

Report

Report Number
2531779-2014-18254
Event Type
Malfunction
Date Received
June 25, 2014
Report Date
June 23, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/23/2014 WITH THE FOLLOWING FINDINGS:DURING A VISUAL INSPECTION OF THE PUMP, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. DURING TESTING, THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL KEYPAD BUTTON CONTACTS. ADDITIONALLY, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED. UNRELATED TO THESE ISSUES, THE KEYPAD COVER WAS FOUND TO BE TORN. ALL OF THE KEYPAD BUTTONS RESPONDED PROPERLY TO USER PRESSES. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A BATTERY COMAPRTMENT CRACK, CONTAMINATION UNDER KEYPAD BUTTON CONTACTS, AND A DISPLAY FAILURE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06/23/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371833 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 10 YR