FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 3895516 · Received June 25, 2014

Report

Report Number
3004753838-2014-04848
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
March 4, 2014
Report Date
June 2, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE RETURNED DEVICE CONFIRMED NO AUDIO OUTPUT.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 AND CLAIMED THAT ON (B)(6) 2014 PATIENT EXPERIENCED NO AUDIBLE ALERTS. DEXCOM HAS ISSUED AN RGA FOR PATIENT TO RETURN THE DEVICE TO BE INVESTIGATED. THERE WAS NO REPORT OF ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
369938 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT20649 5091252

Patients

Seq Age Sex Outcome Treatment
1 3 YR Other