FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 3895505 · Received June 25, 2014

Report

Report Number
3004209178-2014-86142
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. CHECKED O-RINGS/STOPPER GROOVE FOR ANOMALIES. NONE WERE FOUND. RAN BASAL/BOLUS LEAKING TEST. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR LEAKS PAST THE SECOND O-RING.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED LOW RESERVOIR. INSULIN HAS LEAKED INTO THE RESERVOIR COMPARTMENT. THE BLOOD GLUCOSE READING IS 21.6 MMOL/L. CUSTOMER TREATED WITH INSULIN PUMP. CUSTOMER NOTICED THE LEAK WHEN REMOVING THE RESERVOIR FROM THE INSULIN PUMP. THE INSULIN LEAKED PAST THE O-RINGS. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
370229 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A H8962300

Patients

Seq Age Sex Outcome Treatment
1