G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-04843
- Event Type
- Malfunction
- Date Received
- June 25, 2014
- Date of Event
- June 2, 2014
- Report Date
- June 2, 2014
- Manufacturer
- DEXCOM INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE COMPLAINT OF INTERMITTENT VIBRATION COULD NOT BE CONFIRMED. CAPA HAS BEEN INITIATED FOR THIS ISSUE.
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 AND CLAIMED NO DEVICE VIBRATION ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, AT THE TIME OF THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT TO TEST ALERT FUNCTIONALITY AND PATIENT REPORTED ALERTS WERE FUNCTIONAL. DEXCOM HAS ISSUED AN RGA FOR PATIENT TO RETURN THE DEVICE TO BE INVESTIGATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 370487 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS, PRODUCT CODE:MDS | MDS | DEXCOM INC. | MT20649-BLU | 5067283 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other |