VITROS SALI SLIDES
Report
- Report Number
- 1319809-1996-00001
- Event Type
- Malfunction
- Date Received
- August 30, 1996
- Date of Event
- August 6, 1996
- Report Date
- August 6, 1996
- Manufacturer
- JOHNSON &JOHNSON CLINICAL DIAGNOSTICS
- Product Code
- DKJ
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
A TOTAL OF 15 SLIDE CARTRIDGES (270 SLIDES) WERE RETURNED TO CO FOR EVALUATION. THE SLIDES WERE TESTED USING A QUALITY CONTROL FLUID, LOT NUMBER G1272. ALL 270 SLIDES TESTED GAVE RESULTS BETWEEN 12.5 TO 13.2 MG/DL. ALL SLIDES TESTED WERE WITHIN THE EXPECTED QUALITY CONTROL RANGE. PRODUCT PERFORMED WITHIN SPECIFICATIONS. ALLEGED FAILURE COULD NOT BE DUPLICATED.
PT SAMPLE MESURED A SALICYLATE OF <1MG/DL (NEGATIVE) USING CO SALICYLATE SLIDES. RESULT REPORTED TO PHYSICIAN. PHYSICIAN REQUESTED A REPEAT OF THE TEST. A FRESH PT SAMPLE WAS DRAWN AND MEASURED USING THE SAME CARTRIDGE OF SLIDES. RESULT WAS 3.0 MG/DL (THERAPEUTIC RANGE). A NEW CARTRIDGE WAS LOADED AND CALIBRATED ON INSTRUMENT. ORIGINAL SAMPLE AND RE-DRAW WERE RUN USING NEW CARTRIDGE. ORIGINAL SAMPLE GAVE 52MG/DL AND RE-DRAW GAVE 47 MG/DL. BOTH RESULTS ARE IN THE TOXIC RANGE FOR SALICYLATE. PHYSICIAN STATED THAT THERE WERE NO ADVERSE EFFECTS DUE TO THE MISREPORTED RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS SALI SLIDES | IN VITRO DIAGNOSTIC | DKJ | JOHNSON &JOHNSON CLINICAL DIAGNOSTICS | NA | 2206-0009-4400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |