FDA Adverse Event Malfunction Summary report: N

MAVERICK²?

MDR report key: 3895192 · Received June 25, 2014

Report

Report Number
2134265-2014-03645
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
June 2, 2014
Report Date
June 2, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: RETURNED PRODUCT CONSISTED OF A MAVERICK2 BALLOON CATHETER WITH A STOPCOCK ATTACHED TO THE HUB. THE BALLOON WAS TIGHTLY FOLDED. THERE WERE MULTIPLE HYPOTUBE KINKS. THE HYPOTUBE WAS COMPLETELY SEPARATED 78CM FROM THE HUB. THE HYPOTUBE FRACTURE SURFACES WERE OVALED, WHICH SUGGESTS THE DEVICE WAS KINKED PRIOR TO SEPARATION. INSPECTION OF THE REMAINDER OF THE DEVICE PRESENTED NO OTHER DAMAGE OR IRREGULARITIES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE SHAFT BREAK WAS MORE OR LESS THAN 60 CM FROM THE HUB. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 2.00MM X 20MM MAVERICK² CATHETER BALLOON WAS SELECTED TO DILATE THE LESION. DURING ADVANCEMENT, IT WAS NOTICED THAT THE SHAFT WAS BROKEN.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE SHAFT BREAK WAS MORE OR LESS THAN 60 CM FROM THE HUB. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371073 MAVERICK²? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493892820200 16777007

Patients

Seq Age Sex Outcome Treatment
1